Trouble shooting, Probe troubleshooting – Super Systems Gold Probe User Manual

Page 9

Advertising
background image

8

SSi Gold Probe Manual Rev C

TROUBLE SHOOTING

When trouble arises with a furnace control
system, it is important to establish where the
problem is located; the probe, signal
transmission lines, the control instrument, or
the furnace itself. Several simple tests can help
to isolate the problem quickly. It is most
important to first understand the nature of the
fault. Aside from erratic behavior like cycling,
or failure to stabilize at the set point, the most
common symptom is non-conformity of the
work pieces to quality assurance specifications.

To evaluate most faults, the recommended
tools are:

1. a good 3 ½ digit millivolt meter with at

least 10 meg

input impedance and 0 to

1999 mV range,

2. a temperature calibrator and,
3. a simulator to output 0 to 1300 millivolts

at less than 50 megohms output
impedance.


Probe troubleshooting: In order to establish
the source of problems in your installation, first

resist the temptation to remove the Gold
Probe from the furnace.

All of the following

meaningful questions must be answered while
your Gold Probe (or any other carbon sensor) is
in the furnace, at temperature, and exposed to
a normal atmosphere under manual control:

1. Does an Alnor dew point reading (or

shim stock analysis) verify the indicated
value from the probe? If there is reasonable
correlation, the problem is NOT the probe.


2. Are the connections from the T/C extension

wire and sensor cable clean and firmly
attached at the correct probe and control
instrument terminals? Note that the shield
wire in the sensor cable should be
connected to ground at the control
instrument end only!



SYMPTOM POSSIBLE CAUSE
High % C

Low reading due to:

High probe resistance

Cracked zirconia

Dirty reference air

Faulty cable insulation

Instrument calib./ calc.

Air leak to burnoff fitting

Furnace air leak at probe

Oily parts or sooted furnace

Wrong recipe time/temp

Low % C

High Reading due to

Probe plugged with soot

Instrument calib./calc.

Wrong recipe time/ temp

Erratic

Faulty signal due to

Bad sensor connections

Electrical noise source

Radiant tube leak

Bad Endo

Mixing valve setting

Instrument setting

Sooted fce

Endo not cracked (temp too

Low in generator or
Catalyst inactive)

3. Is the control instrument CO or H

2

factor set to

the appropriate value? This “factor” is referred
to by various manufacturers as Zone Factor,
Process Factor, Gas, Furnace Factor, CO
Factor, Calibration Factor, etc. This factor may
require adjustment in order to make the
calculated %C or dew point agree with other
measurements.


4. Do the actual Gold Probe temperature and O

2

mV signal, as measure by the temperature
calibrator, and digital voltmeter, agree with the
displayed values on the control instrument? If
not, and instrument calibration problem is
likely.

Advertising