Watson-Marlow 520Du User Manual

Page 12

Advertising
background image

Watson-Marlow 520Du User Manual

12

Pump specifications

Control range (turndown ratio)

0.1-220rpm (2 200:1)

Supply voltage/frequency

100-120V/200-240V 50/60Hz 1ph

Maximum voltage fluctuation

±10% of nominal voltage.

A well regulated electrical mains supply

is required along with cable connections

to the best practice of noise immunity

Installation category
(overvoltage category)

II

Power consumption

135VA

Full load current

<0.6A at 230V; <1.25A at 115V

Eprom version

Accessible through pump software

Enclosure rating

IP31 to BS EN 60529. Equivalent to

NEMA 2, suiable for indoor use.

Protected against dripping water and

falling dirt. May be wiped with a damp

cloth, but should not be immersed

Pumphead options

520R, 501RL, 313, 314, 505L,

505BA, 505CA, 314MC, 318MC

Operating temperature range

5C to 40C, 41F to 104F

Storage temperature range

-40C to 70C, -40F to 158F

Maximum altitude

2,000m, 6,560ft

Humidity (non-condensing)

80% up to 31C, 88F, decreasing linearly

to 50% at 40C, 104F

Pollution degree

2

Noise

<70dB(A) at 1m

* Protect from prolonged UV exposure.

Note: 520 drive models are c ETL us listed. Cert to std CAN/CSA C22.2 No 1010-
92. Conforms to std UL 61010A-1 April 30, 2002.
Note: 520 drive models have been tested in accordance with BS EN 61000-6-
2:2001 (EN 61000-4-4) Fast Transient and Burst Tests to Industrial limits - ie:
Level 3 : 2kV.

Advertising
This manual is related to the following products: