5.5 List of functions

1. Volume

2. Tube dimension

20. Operator number

3. Velocity

21. Batch number

4. Acceleration/deceleration

24. Print status

5. Reversing (back suction)

29. Print parameters

6. Batch size

31. Save program

7. Delay

32. Load program

8. Completed fills

33. Delete program

9. Specific gravity

34. Print programs

10. Output rate

46. Set language

15. Input mode

47. Printer set-up

18. Timer 2

72. Volume format

19. Timer 3

80. Reset memory

5.6 Description of functions

The individual functions will be described in the following:

1. Volume
Value: Choice of ml and grams.
Range: Required volume from 1 to 9999.9 ml.

2. Tubes
Value: Inside diameter (i.d.) in mm
Range: 3.0 - 5.0 - 6.5 - 8.5 - 10.0 - 12.5
The tubes can be measured with the supplied tube gauge.

If the tube is changed, a new calibration must always be made.

3. Velocity
Value: Pump revolutions per minute (rpm)
Range: 30 - 250 rpm

If the velocity is changed, a recalibration should be made.

4. Acceleration/deceleration
Value: An integral number
Range: 1 - 100
The acceleration and the deceleration will always be the same. The lowest value (1) will give the
lowest acceleration, and the highest value (100) will give the fastest acceleration.

If the acceleration is changed, a recalibration should be made.