8 Alarm and warnings

8.1 Alarms handling


When MC100 is in Alarm state, the only possible action is to reset the alarm. Only applies if the alarm
causes are removed, otherwise the alarm will re-appear at the next attempt.

Alarms are cleared one-by-one, starting top down – meaning lowest number first.

8.1.1 Module alarms


Alarm no.

Alarm description

Action

0

No pending alarms

1

No pumps connected during
power up

Check connections /power to pumps

2

Connection to one or more
pumps lost

Check connections /power to missing pump

3

Connected pumps not equal to
last time

Configuration changed – Investigation or new
initialization must be done

4

Attempt to access not
connected pump

5

Reverse direction not available

Pump command Bit 5: Direction
set for one direction pump

6

Unstable FlexNet: Check
pumps

7

7

FRAM storage failure: Restart
or Reset


8

Mode change not completed:
check pumps

Changing mode was not possible, check pump
versions, if change was to parallel or serial mode

9

Missing or defect Anybus
CompactCom Module

No Alarm string available, because module not working

10

Reserved

11

Run length limit

Run less than 5 or more than 16777215 steps

12 -17

Reserved

NN

FB module not Ready

8

Internal MC100 module check, possible replacement of
module.

8.1.2 Pump alarms

20

Dispense not finished

Dispensing has stopped before finishing.

21

Tube bridge off at dispense start

Dispensing or pumping cannot start

22

Tube bridge off while dispensing

Dispense not finished

23

Tube bridge off at pumping start

Trying to start pumping with tube bridge off

24

Tube bridge off while pumping

Removing tube bridge while pumping

7

Expected to be implemented in a later version

8

Alarm is only displayed with the 2 LED’s on MFSC module – both flashes RED.