Remote operation over the rs-232-c/v.24, Data interface – HEIDENHAIN ND 231 B v.2 User Manual
Page 51
51
Remote
O
peration
over
the
R
S-232-C/V.24
Data
Interface
Remote Operation over the RS-232-C/V.24 Interface
You can operate the display unit remotely over the RS-232-C/
V.24 data interface. The following commands are available on
the ND 231 B:
Format:
<ESC>TXXXX<CR> Pressed key
<ESC>AXXXX<CR> Output of screen contents
<ESC>FXXXX<CR> Execute function
<ESC>SXXXX<CR> Special function
Sequence of commands
Meaning
<ESC>T0000<CR>
‘0‘ key
<ESC>T0001<CR>
‘1‘ key
<ESC>T0002<CR>
‘2‘ key
<ESC>T0003<CR>
‘3‘ key
<ESC>T0004<CR>
‘4‘ key
<ESC>T0005<CR>
‘5‘ key
<ESC>T0006<CR>
‘6‘ key
<ESC>T0007<CR>
‘7‘ key
<ESC>T0008<CR>
‘8‘ key
<ESC>T0009<CR>
‘9‘ key
<ESC>T0100<CR>
‘CL‘ key
<ESC>T0101<CR>
‘-‘ key
<ESC>T0102<CR>
‘.‘ key
<ESC>T0104<CR>
‘ENT‘ key
<ESC>T0105<CR>
‘MOD‘ key
<ESC>T0107<CR>
‘1/2‘ (datum) key
Sequence of commands
Meaning
<ESC>T1000<CR>
‘CE+0‘ keys
<ESC>T1001<CR>
‘CE+1‘ keys
<ESC>T1002<CR>
‘CE+2‘ keys
<ESC>T1003<CR>
‘CE+3‘ keys
<ESC>T1004<CR>
‘CE+4‘ keys
<ESC>T1005<CR>
‘CE+5‘ keys
<ESC>T1006<CR>
‘CE+6‘ keys
<ESC>T1007<CR>
‘CE+7‘ keys
<ESC>T1008<CR>
‘CE+8‘ keys
<ESC>T1009<CR>
‘CE+9‘ keys
<ESC>A0000<CR>
Output of model designation
<ESC>A0100<CR>
Output of 14-segment display
<ESC>A0200<CR>
Output of current value
<ESC>A0301<CR>
Output of error text
<ESC>A0400<CR>
Output of software number
<ESC>A0900<CR>
Output of status indicators
<ESC>F0000<CR>
REF function
<ESC>F0002<CR>
<ESC>S0000<CR>
Counter RESET
<ESC>S0001<CR>
Lock keypad
<ESC>S0002<CR>
Release keypad