- obviously incorrect measurements occur.
- measurements with the unit are no longer possible.
- the permissible limiting values were exceed.

4. Safety regulations

In order to exclude incorrect operation of the unit, its use assumes
that the user is aware of and complies with the usual safety regu-
lations.
- the batteries may only be changed outside the Ex-area.
- it is not allowed to take additional batteries into the Ex-area.
- only type approved batteries may be used.
- only sensors which have been approved by ecom may be used.
- It must be ensured that the unit is not taken into Zone 0.
- measurements on electrically live parts are not allowed.
- observe the measuring range of the temperature sensors.

O

Overheating will damage the sensors.

- avoid use in aggressive acids or alkalis.
- the unit may not be opened in Ex-areas.

4.1 Special conditions for use in Zone 0

The measurement of temperatures with the Ex-Pt 720 within
Zone 0 is only permitted if the following conditions are complied
with:

- a stainless-steel sensor Type 0628 2232 is used.
- the Ex-Pt 720 itself may not be taken into Zone 0.
- It must be absolutely ensured that the unit does not find its way

accidentally into Zone 0. In order to guarantee this, the carrying

st

strap of the unit must be attached additionally to the belt of the

user or to a loop attached to his belt. The length of the carrying

st

strap must be so adjusted that when the unit hangs freely on the

carrying strap, the Ex-Pt 720 maintains a spacing of 50cm to the

b

boundary area of Zone 0.

- Should the unit be used in the optionally available leather case,

t˝˝

the carrying strap of the unit must still comply with the conditi-

ti˝

ons for attachment to the user’s body.

5. Ex-Data

EC-test approval-no.: TÜV 01 ATEX 1757 X
Ex-designation:

II 2(1) G EEx ia IIC T4

Approved for Zone 1, equipment group II, gas group C explosions-
endangered gases, vapours or fog, temperature class T4.

Only the sensors specified in the approval certificate may be used.
Sensor no. 0628 2232 may be used in Zone 0 !

6. Technical Details

Ambient temperature Ta:

-10 ... +50°C

Storage temperature:

-20 ... +70°C

Batteries:

9V-Block, 6LR61 as per IEC
(only type tested batteries)

Supplier

Type

Varta

Alkaline No.4822

Varta

Alkaline Universal No.4022

Varta

Alkaline Electric Power No.8022

Duracell

Alkaline

Duracell

Alkaline Ultra

Duracell

Professional Alkaline Battery Procell

Everady
(Ralston Energy Systems SA)

Alkaline Energizer

Panasonic

Alkaline Power Line Industrial Battery

Daimon

Alkaline