Ecom Instruments Ex-Pt 720 User Manual

Page 5

Advertising
background image

22

23

Pressing the >Hold-pushbutton< switches between °C and °F.
Should no pushbutton be pressed for 5 seconds after a changeo-
ver, then the new measuring value will be accepted.

7.2.5 Switching-off

Switch the unit of with the >I/O pushbutton< . The display is no
longer active.

7.3 Battery change

Should the <BAT> symbol be displayed constantly, then the bat-
tery will need to be changed within the next 1-2 hours. The unit
switches itself automatically off should the battery voltage be too
low.
Open the back of the unit by unscrewing it, remove the exhausted
battery and insert a new one (observe polarity).
- while changing the battery, the cover may not be separated from

t˝˝

the unit top shell.

- only batteries of Type 6LR61 as per IEC may be used. The use of

al

alternative types of batteries is strictly forbidden.

- batteries may only be changed outside the Ex-area.
- the taking of additional batteries into the Ex-area is not allowed.

7.4 Temperature probes

7.4.1 Standard probes

Immersion / insertion probe

0628 1232:

measurement range:-50 ... 400°C
accuracy:

class A

Ex-Zones:

2, 1

probe dimensions: 110mm x 4mm (w x ø)
sensor dimensions: 30mm x 3,2mm (w x ø)
response time (99%): 12 Sec.

Surface probe

0628 1932:

measurement range:-50 ... 400°C
accuracy:

class B

Ex-Zones:

2, 1

probe dimensions: 110mm x 4mm (w x ø)
sensor dimensions: 3,2mm (ø)
response time (99%):40 Sec.

Stainless-steel probe

0628 2232:

measurement range:-50 ... 400°C
accuracy:

class A

Ex-Zones:

2, 1, 0

probe dimensions: 125mm x 4mm (w x ø)
sensor dimensions: 15 x 3mm (w x ø)
response time (99%):10 Sec

7.4.2 Special probes

Immersion / insertion probe

0600 9999

measurement range:-50 ... 400°C

0628 1232:

accuracy:

class A

Ex-Zones:

2, 1

probe dimensions: >50mm x >4mm (w x ø)
sensor dimensions: >30mm x >3,2mm (w x ø)
response time (99%):12 Sec.

Surface probe

0600 9999

measurement range:-50 ... 400°C

0628 1932:

accuracy:

class B

Ex-Zones:

2, 1

probe dimensions: >50 mm x >4mm (w x ø)
sensor dimensions: >9 mm (ø)
response time (99%):40 Sec.