If the [E-##] or [Err] error messages are displayed, the unit
will not function and must be returned to the factory for
repair. (See section on Warranty for return instruction).

The following charts summarize the fault condition, and the
associated indication.

5.2

Model DX 501

[LO] Target < -50 °F

(-45 °C)

Contact
Exergen

Model DX 1001

[LO] Target <0 °F (-

18 °C)

Use lower DX

Model DX 1201 [LO] Target <186 °F

(86 °C)

Use lower DX

Model DX 1601 [LO] Target <186 °F

(86 °C)

Use DX 501
or DX 1001

Low Target Temperature

Condition

Error

Code

Remedy

Measurement
Fault

High Ambient
Temperature

[HI A] Ambient > 122

°F (50 °C)

Move to
alternate
location

Low Ambient
Temperature

[LO A] Ambient < 32

°F (0 °C)

Move to
alternate
location

Low Battery

[N/A]

Battery power
low

Replace
battery

Dead Battery

[----]

Dead battery

Replace
battery

Other Errors

Condition

Error

Code

Remedy

Measurement
Fault

Processing
Error

[E-##]

or [Err}

Internal fault

Repeat
measurement,
if error
remains,
contact
Exergen

8

Model DX 501

[HI]

Target >550°F
(287 °C)

Use higher
DX

Model DX 1001

[HI]

Target >1000
°F (538 °C)

Use higher
DX

Model DX 1201

[HI]

Target >1207
°F (653 °C)

Use higher
DX

Model DX 1601

[HI]

Target >1600
°F (871 °C)

Contact
Exergen

High Target Temperature

Condition

Error

Code

Remedy

Measurement
Fault