Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©

Page 92

www.PressureSystems.com

SET/DO OPERATING OPTIONS/FUNCTIONS (Command 'w')


Purpose:

Change a module’s default operating option settings, or invoke special internal
operations or functions.

Command

wii[dd[ eeee]]

‘w’ is the command letter.
ii’ is the index field.
dd’ is an [optional] hex datum.
eeee’ is an [optional] extra datum
with leading space character.

Response

A

‘A’ is the acknowledge letter.


Description: The index field (ii) contains two hex digits that identify the specific option to be

set or function to be performed. The datum field (dd), when present, contains 2
hex digits. A few indexes also require an extra datum field (eeee). Valid
options/functions are listed in the table below (-- marks a missing datum field in
its column, and fourth column shows any 'q' command index that reads same
option):

ii

dd

Description

'q' read

index

00 --

Execute Internal Self Test.

01 --

Update Internal Thermal Coefficients.

02-06

--

Reserved for factory use

07 --

Store Most Operating Options in nonvolatile flash
memory. See *** Note at end of table.

08 --

Store Current Offset Terms in transducers’ non-volatile
memories. See *** Note at end of table.

09 --

Store Current Gain Terms in transducers’ non-volatile
memories. See *** Note at end of table.

0A 01-10

(1-16)

Set Number of Channels in Module (default =16).

0B

00

01

Enable Automatic Shifting of Calibration Valve during
Calculate and Set Offsets ('h') command (default).
Disable Automatic Shifting of Calibration Valve in 'h'.
User will manually control calibration value.