Emerson Process Management 430 DO User Manual

Page 6

Advertising
background image

2

MODEL 430 DO

SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

1.3 AVAILABLE MODELS
The Model 430-01 and -02 Sensor is a flow through

assembly. The sensor is supplied in a 1-1/2 inch PVC

tee for installation into a process line.
The Model 430-03 Submersible Sensor is for use in

those applications where the required flow velocity can

be maintained without the use of an agitator. When the

measurement is being made in a stagnant pond or

other media where the minimum flow rate of 1.5 FPS

cannot be obtained, the Model 430-04 must be select-

ed. This model is designed for submersible use and is

provided with its own agitator to ensure adequate flow

velocity at the membrane.
The Models 430-03 and 04 may be provided with an

optional mounting bracket assembly. This bracket is

constructed of light weight aluminum and can be

adjusted to satisfy virtually any aeration basin mount-

ing requirement. Once installed, the probe can be

removed without the use of tools. A single locking pin

holds the assembly rigidly in place but allows it to be

easily removed for inspection or cleaning.

1.4 SPECIFICATIONS
Process Connection:

430-00/01 1-1/2 inch socket fitting

Wetted Materials: HOUSING - PVC

MEMBRANE - Teflon

PRESSURE COMPENSATOR - Neoprene

Temperature Compensation:

0 to 50°C (32 to 122°F) Automatic

Sample Pressure: 0-50 psig (345 kPa abs)

Submersible to 200 ft. (61 m).

Agitator Power Requirement: Model 430-04 only)

115 Vac, ±10 60 Hz (0.2 amp)

Cable Length (standard):

6.1 meters (20 ft) on sensor and agitator

Maximum Cable Length:

305 meters (1000 ft)

Operating Time Between Electrolyte Recharge:

Approximately 12 months or longer, depending upon

application

Sample Flow Requirements for Models 430-01-02-03:

1.5 feet per seconds (0.5 m/s)

FIGURE 1-1 Flow Through Assemblies

DWG. NO.

REV.

40043019

A

MILLIMETER

INCH

Advertising