Emerson Process Management Micro Motion 3500 Transmitter User Manual

Page 8

Advertising
background image

7

STEP 4.

Connecting input and output wiring

Connect input and output wiring to the appropriate terminals on the
input/output wiring connector, which is the far right connector. Refer to
Table 1 and to the card that is inserted into the sleeve on the top panel
(shown in Figure 4).

Use 24 to 16 AWG (0,25 to 1,5 mm

2

) twisted-pair shielded wire.

Ground the shields at a single point only.

Figure 4.

Input/output wiring terminal card

STEP 5.

Connecting the Model 3500 to the sensor

To connect the Model 3500 transmitter to a Micro Motion sensor, follow
the steps below. If you are installing the Model 3300 applications
peripheral, this step is not required.

Table 1.

Input/output wiring terminals

Terminal number

Designation

c 2+

a 2 –

Primary 4–20 mA output

c 4 +

a 4 –

Secondary 4–20 mA output

c 6 +

a 6 –

Frequency input

c 8 +

a 8 –

Discrete input 1

c 10 +

a 10 –

Discrete input 2

c 12 +

a 12 –

Frequency output

c 14 +

a 14 –

Discrete output 1

c 16 +

a 16 –

Discrete output 2

c 18 +

a 18 –

Discrete output 3

c 32 (B line)

a 32 (A line)

RS-485 output

Advertising