Warning, Caution – Precision Medical PM5400 Series User Manual

Page 4

Advertising
background image

3

H

ELI

O

2

BLENDEr

HELIUM-OxygEn

WARNING

• Only trained, qualified medical personnel under the direct supervision of a

licensed physician should operate the

H

ELI

O

2

Blender.

• Use this

H

ELI

O

2

Blender only for its Intended Use as described in this

manual.

• Confirm prescribed dose before administering to patient. Monitor on a

frequent basis. Continuous Monitoring with an alarmed Oxygen Monitor

/ Analyzer is recommended.

• The

H

ELI

O

2

Blender shall be serviced by a qualified service technician.

• Always follow ANSI and CGA standards for Medical Gas Products,

Flowmeters and Oxygen Handling.

• The clinician must verify the Heliox tank concentration prior to utilizing

the Blender.

• An alarmed Oxygen Monitor / Analyzer must be used to verify oxygen

concentration.

• Accuracy of oxygen concentration will be affected if bleed is not activated

at flow settings below 15 l/min for the High Flow Blender, and 3 l/min for

the Low Flow Blender.

• dO NOT obstruct the alarm.

• dO NOT use Blender when alarm is sounding.

• dO NOT use oil in or around the Blender.

• dO NOT occlude or obstruct the bleed port on the auxiliary outlet of

the Blender.

• dO NOT use near any type of flame or flammable/explosive

substances, vapors or atmosphere.

• Oxygen Concentration dial does not rotate 360 degrees. Rotating the

dial beyond the endpoint settings will damage the Blender.

CAUTION

• Turn off gas supplies when

H

ELI

O

2

Blender is not in use.

• Store the

H

ELI

O

2

Blender in a clean, dry area when not in use.

• The

H

ELI

O

2

Blender contains magnetic, ferrous material that may

affect the results of an mRI.

• Ensure all connections are tight and leak free.

• Avoid excessive pressure surges greater than 100 psi (6.9 bar) when

pressuring the Blender inlets.

• DO NOT steam autoclave.

• DO NOT immerse

H

ELI

O

2

Blender into any liquid.

• DO NOT gas sterilize with (EtO) Ethylene Oxide.

• DO NOT use if dirt or contaminants are present on or around the

Blender or connecting devices.

• DO NOT smoke in an area where oxygen is being administered.

• DO NOT clean with aromatic hydrocarbons.

• Inlet pressure of device used in conjunction with the

H

ELI

O

2

Blender must

match inlet pressure of

F

I

O

2

Blender.

• When using a bottled high pressure gas source, always use a pressure

reducing regulator set within 30-75 psi (2.1-5.2 bar).