Specifications – Precision Medical PM5400 Series User Manual

Page 5

Advertising
background image

4

H

ELI

O

2

BLENDEr

HELIUM-OxygEn

SPECIFICATIONS

High Flow

Low Flow

Model

PM5580: 80/20 heliox

PM5570: 70/30 heliox

PM5480: 80/20 heliox

PM5470: 70/30 heliox

Primary Outlet

Flow Range

15 - 120 l/min

3 - 30 l/min

With both supply pressures at 50 psi (3.4 bar)

with BLEED Closed

Auxiliary Outlet

Flow Range

2 - 100 l/min

0 - 30 l/min

With both supply pressures at 50 psi (3.4 bar)

with BLEED Open

Bleed Flow @

100%

F

I

O

2

13 l/min or less

at 50 psi (3.4 bar)

3 l/min or less

at 50 psi (3.4 bar)

Maximum

Combined

Flow (All Outlets)

120 l/min

30 l/min

Bypass Flow

(Loss of Helium or

Oxygen supply)

>

85 l/min

>

45 l/min

Bypass Alarm

Activation

50 psi

(3.45 bar)

60 psi

(4.14 bar)

50 psi

(3.45 bar)

60 psi

(4.14 bar)

13-25 psi

16-24 psi

18-22 psi

16-24 psi

0.9-1.7 bar 1.1-1.65 bar 1.2-1.5 bar 1.1-1.65 bar

Alarm Reset:

When pressure differential is 6 psi (0.4 bar)

or less.

Alarm Sound Level:

to 80 db at 1 ft (0.3 m)

Oxygen Concentration

Adjustment Range:

PM5480

PM5580

20 -

100%

PM5470

PM5570 30 - 100%

Gas Supply Pressure:

30 - 75 psi (2.1 - 5.2 bar) Heliox & Oxygen

within 10 psi (0.69 bar) of each other

Mixed Gas Stability:

±1% Oxygen

Connection Types:

DISS Type - Heliox & Oxygen Inlets & Outlets

and / or NIST Type - Heliox & Oxygen Inlets

Note: All flow-rate values are as measured from an Oxygen flowmeter (uncorrected).