12

TROUBLESHOOTING

If the Conserver fails to function, consult the Troubleshooting Guide

below. If problem cannot be solved, consult your Provider.

TROUBLESHOOTING GUIDE

Problem

Probable Cause

Remedy

No flow

1. Cylinder valve

closed

2. Conserver in “OFF”

position

3. Cylinder empty
4. Dial not set at

proper position

5. Conserver not

sensing breath

1. Open the valve on

cylinder

2. Set to prescribed

setting

3. Replace cylinder
4. Set dial so indicator

points to a setting

5. Check position of

cannula in nose,
DO NOT breathe

through mouth

Leaking at

cylinder

connection

1. Missing or defective

yoke seal

2. Defective cylinder

valve

1. Replace with

Precision Medical

yoke seal

2. Contact your

Provider