8

INSTALLATION

WARNING

• Read this User Manual before installing or operating the

Oxygen Conserving Regulator.

• Inspect the Conserver and cylinder valve to ensure they are

free of oils, greases or other contaminants.

• DO NOT direct flow of oxygen at any person, or flammable

material when cracking open the cylinder.

• DO NOT use without yoke seal.

• DO NOT use plastic yoke seals.

• Use Precision Medical, Inc. yoke seal made of viton and brass.

• The yoke seal supplied with the Conserver is reusable.

CAUTION

• Inspect the Oxygen Conserving Regulator for visual damage

before use, DO NOT USE if damaged.

• DO NOT use tools to tighten T-handle, this can lead to over

tightening and will cause damage to the Conserver.

• Be sure all connections are tight and leak free.
• DO NOT use liquid leak detector to test for leaks.

NOTE: For operational instructions on how to operate the

cylinder consult your Provider.

1. Position the cylinder so the oxygen cylinder valve outlet is pointing

away from the user and any other person(s).

2. Before installing the Conserver, “crack” cylinder valve by opening the

valve slightly to remove any foreign particles and then close valve.

3. Be sure the Conserver is in the “OFF” position. Align “OFF” with the

indicating pointer .

4. Be sure the CGA 870 connection high pressure inlet has a yoke seal.
5. Align Index Pins with Index Pin Alignment Holes in Post Valve.
6. Hand tighten T-handle until snug.