Waveform Invert

Press CH1Invert to turn on or off waveform invert. When waveform invert is
turned off, the waveform display is normal; when waveform invert is turned on, the

waveform voltage values are inverted (as shown in the figures below).

(a) "Invert" Off

(b) "Invert" On

Figure 2-1 Waveform Invert

Vertical Scale

Vertical scale refers to the voltage value per grid in the vertical direction on the

screen and is usually expressed as V/div.


Press CH1 and rotate VERTICAL SCALE to adjust the vertical scale (clockwise
to reduce the scale and counterclockwise to increase). The size of the displayed

waveform will be changed accordingly. The scale information (as shown in the figure

below) in the channel label at the bottom of the screen will change accordingly

during the adjustment. The adjustable range of the vertical scale is related to the

probe ratio currently set. By default, the probe ratio is 10X and the adjustable range

of the vertical scale is from 10 mV/div to 100 V/div.


The vertical scale can be adjusted in "Coarse" or "Fine" mode. Press CH1

Volts/Div to switch the adjustment mode.
 Coarse adjustment (take counterclockwise as an example): set the vertical scale

in 1-2-5 step namely 10 mV/div, 20 mV/div, 50 mV/div, 100 mV/div…100 V/div.

 Fine adjustment: further adjust the vertical scale within a relatively smaller

range to improve vertical resolution. If the amplitude of the input waveform is a

little bit greater than the full scale under the current scale and the amplitude

would be a little bit lower if the next scale is used, fine adjustment can be used

to improve the amplitude of waveform display to view signal details.

Note:

You can also press VERTICAL SCALE to quickly switch between "Coarse"