Troubleshooting


1
When pressing the power switch, the Multimeter has

blank screen with nothing displaying:

(1) Check if the power is correctly connected.

(2) Check if the main power switch on the rear panel has been

turned on.

(3) Check if the safety fuse has been blown, replace it if necessary.

(4) Restart the instrument.

(5) If it still can’t work properly, please contact

RIGOL for help.

2 When connecting a current signal, the reading has not

change:

(1) Check if the test lead is correctly connected to current jack or

the LO jack.

(2) Check if the safety fuse in the current location on the rear panel

has blown.

(3) Check if the measure location has switched to the DCI or ACI

place correctly.

(4) Check whether the input is ACI but the shelves location is DCI.

3 When connecting a DC power signal, the reading

display is abnormality:

(1) Check if the test lead is correctly connected with the current

jack or the LO jack.

(2) Check if the safety fuse in the current location on the back

panel has been blown.

(3) Check the measure location has switched to the DCI or DCV

place correctly.

(4) Check whether the input is DCI but the shelves location is ACI.