Amprobe ACD-16-PRO Clamp-Meters User Manual

Page 14

Advertising
background image

13

True RMS Models ACD-16TRMS & ACD-40PQ ACV & ACA clamp-on accuracies are

specified from 5% to 100% of range or otherwise specified. Maximum Crest Factor are as
specified below, and with frequency spectrums, besides fundamentals, fall within the meter
specified AC bandwidth for non-sinusoidal waveforms. Fundamentals are specified at 50Hz
and 60Hz.


AC Voltage
RANGE Accuracy
50Hz / 60Hz
600.0V

1.0% + 5d

45Hz ~ 500Hz
600.0V

1.5% + 5d

500Hz ~ 3.1kHz (model 132 & 135 only)
600.0V

2.5% + 5d

CMRR : >60dB @ DC to 60Hz, Rs=1k

Input Impedance: 2M

Ω, 30pF nominal

True RMS models Crest Factor:

< 2.3 : 1 at full scale & < 4.6 : 1 at half scale


DC Voltage
RANGE

Accuracy

600.0V 0.5%

+

5d

NMRR : >50dB @ 50/60Hz
CMRR : >120dB @ DC, 50/60Hz, Rs=1k

Input Impedance: 2M

Ω, 30pF nominal


THD%-R

1)

(model 135 only)

RANGE Harmonic

order

Accuracy

2)

Fundamental

1.5% of Reading + 6d

2nd ~ 3rd

5.0% of Reading + 6d

4th ~ 10th

2.5% of Reading + 6d


0.0% ~99.9%

11th ~ 51st

2.0% of Reading + 6d

1)

THD-R is defined as: (Total Harmonic RMS / Total RMS) x 100%

2)

Specified accuracy @ ACA fundamental > 5A ; ACV fundamental > 50V


Ohms
RANGE Accuracy

999.9

1.0% + 6d

Open Circuit Voltage : 0.4VDC typical

Advertising