Amprobe DM9C Digital-Multimeter User Manual

Page 9

Advertising
background image

Diode Test

Prior to any diode test, it must be ensured, that the diode to be tested does not

have live voltage across it. Failure to comply with this prescription can lead to

dangerous user injuries or cause instrument damage. Additionally, foreign

voltages falsify the measurement result.

Resistors and semiconductor paths in parallel to the diode cause falsified

measurement results.

1) Position measurement function selection switch to

measurement range.

2) Connect the black test lead to the COM socket and the red test lead to

the

socket.

3) Connect test leads to UUT.

4) Read the measurement result displayed on the screen.

Current Measurement

Ensure that the measurement circuit is not live when connecting the

measurement instrument.

The instruments may only be used in current circuits protected at 16 A up to a

nominal voltage of 600 V.

The nominal cross section of connecting line has to be respected and a safe

connection has to be ensured.

After instrument fuse tripping eliminate the cause for the tripping prior to fuse

replacement.

7

2526044_Rev2_dm9c.qxd 10/26/06 9:50 AM Page 7