General Specification

6

Caution:

Make sure the power cord is the correct type that is required in your area.
This DMS-H1616 has a universal power supply that allows operation in either
100-120V AC or 200-240V AC voltage areas (no user adjustment is required).

Use the proper power cord with correct attachment plug type. If the power source
is 120 V AC, use a power cord which is a Hospital Grade Power Cord with NEMA
5-15 style plug, labeled for 125 volts AC with UL and C-UL approvals. If the power
source is a 240 V AC supply, use the tandem (T blade) type attachment plug with
ground conductor power cord that meets the respective European country's safety
regulations.

The hospital-grade plug for medical products intended for use in Denmark has
DEMKO approval and is rated 13 amps at 250Vac. Plug is recommended for use
in medical applications and specifications are being added to the standard SB
107-2-D1. Plug mates with maker's Danish hospital-grade socket. Hospital sockets
have slightly different shaped openings allowing only the hospital plug, not the
standard Danish plug, to be inserted, to protect the ac circuit in specific medical
settings.

EEC Safety Compliance:
This display unit meets the requirements of EN-60601-1 so as to
conform to the Medical Device Directive 93/42/EEC (general safety

information).

This monitor complies to the above standards only when used with the supplied
medical grade power supply. RPS-160 Series (Mean well)