Cleaning the instrument – Exergen Temporal Scanner 2000C User Manual

Page 12

Advertising
background image

Cleaning the Instrument

The TemporalScanner is an optical instrument. Like a camera or eye glasses,
a dirty lens will distort the view. If the thermometer is unable to see the heat
clearly, it will be unable to measure it accurately, resulting in low readings.

z

Probe lens and cone should be shiny clean, if
not, wipe with a small cloth or swab moistened
with alcohol.

z

Hold upside-down to prevent excess moisture
from entering the sensor area. It will not harm
the sensor, but if it becomes too wet, you will
be unable to take a temperature until it dries.

z

Thermometer case can be cleaned with any
hospital approved disinfectant, alcohol, even
bleach solutions. Avoid gritty, abrasive cleaners

as they can scratch the thermometer.

z

Do not hold the TemporalScanner under the
faucet or submerge in water. It is not water-
proof.

12

tat2000instrev6.qxd 2/6/2006 9:35 AM Page 12

Advertising