Extech Instruments 470 User Manual

Page 5

Advertising
background image

V5.0 5/04

5

Specifications

Function Range Resolution

Accuracy

400mV 0.1mV

±(0.3% reading + 2 digits)

4V 0.001V
40V 0.01V
400V 0.1V

±(0.5% reading + 2 digits)

DC
Voltage

1000V 1V

±(0.8% reading + 3 digits)

50 to 400Hz 400Hz to 1kHz

400mV 0.1mV

±(1.5% reading

+ 15 digits)

±(2.5% reading

+ 15digits)

4V 0.001V
40V 0.01V
400V 0.1V

±(1.5% reading

+ 6 digits)

±(2.5% reading

+ 8 digits)

AC
Voltage

750V 1V

±(1.8% reading

+ 6 digits)

±(3% reading +

8 digits)

400

µA 0.1µA

4000

µA 1µA

40mA 0.01mA
400mA 0.1mA

±(1.5% reading + 3 digits)

4A 0.001A

DC
Current

20A 0.01A

±(2.5% reading + 5 digits)

50 to 400Hz 400Hz to 1KHz

400

µA 0.1µA

4000

µA 1µA

40mA 0.01mA
400mA 0.1mA

±(1.8% reading
+ 8 digits)

±(3.0% reading
+ 7 digits)

4A 0.001A

AC
Current

20A 0.01A

±(3.0% reading
+ 8 digits)

±(3.5% reading
+ 10 digits)

NOTE:: Accuracy is stated at 65

o

F to 83

o

F (18

o

C to 28

o

C) and less

than 75% RH.

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176