Extech Instruments 470 User Manual

Page 6

Advertising
background image

V5.0 5/04

6

Function Range Resolution

Accuracy

400

Ω 0.1Ω

±(0.8% reading + 4 digits)

4k

Ω 0.001kΩ

±(0.8% reading + 2 digits)

40k

Ω 0.01kΩ

400k

Ω 0.1kΩ

4M

Ω 0.001MΩ

±(1.0% reading + 2 digits)

Resistance

40M

Ω 0.01MΩ

±(3.0% reading + 5 digits)

40nF 0.01nF

±(5.0% reading + 7 digits)

400nF 0.1nF

±(3.0% reading + 5 digits)

4

µF 0.001µF

40

µF 0.01µF

±(3.5% reading + 5 digits)

Capacitance

100

µF 0.1µF

±(5.0% reading + 5 digits)

5.000Hz 0.001Hz
50.00Hz 0.01Hz

±(1.5% reading + 5 digits)

500.0Hz 0.1Hz
5.000kHz 0.001kHz
50.00kHz 0.01kHz
500.0kHz 0.1kHz

±(1.2% reading + 2 digits)

5.000MHz 0.001MHz
10.00MHz 0.01MHz

±(1.5% reading + 4 digits)

Frequency

Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms
min @ 20% to 80% duty cycle and > 100kHz

.

0.1 to 99.9% 0.1%

±(1.2% reading + 2 digits)

Duty Cycle

Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz

-4 to 1382

°F 1°F

Temp
(type-K)

-20 to 750

°C 1°C

±(3.0% reading + 3 digits)
(probe accuracy not included)

-58 to 518

°F 1°F

Temp
(IR)

-50 to270

°C 1°C

±2.0% reading or ±2°C, ± 4 °F

Note: Accuracy specifications consist of two elements:
• (% reading) – This is the accuracy of the measurement circuit.
• (+ digits) – This is the accuracy of the analog to digital converter.

Advertising