Rewinding your pump, Starting on insulin 47 – Energizer 715 User Manual

Page 69

Advertising
background image

Starting on insulin 47

Rewinding your pump

Before you continue, make sure the pump is NOT connected to your body.

1

If you removed your reservoir and are replacing it, go to the REWIND screen.

Main > Prime > Rewind

2

In the REWIND screen, press

ACT

to start the

rewind process. The REWINDING screen will
appear while the pump rewinds.

3

After the pump rewinds the PREPARING TO PRIME
screen will appear.

NOTE -

If you press

ESC

or the pump returns

to the HOME screen the PRIMING STOPPED
screen will appear which guides you back to the
priming screens. Once you start a rewind, you
cannot cancel it.

If you are practicing, do these steps:

a.

Do NOT insert the reservoir in your pump.
Make sure the shipping cap is installed in the reservoir compartment.

b.

Continue with the manual prime instructions described in “Manual prime” on page 49.

If you are not practicing, continue to “Inserting the reservoir in your pump” on page 48.

WARNING: Make sure the infusion set is disconnected from your body before you rewind or

prime the pump. Never insert the reservoir into the pump while the tubing is
connected to your body. Doing so could result in an accidental infusion of
insulin.

Disconnect from Body

Press ACT to Rewind

Wait for notification

REWIND

REWINDING

ACT

PRIMING STOPPED

Continue Priming

Rewind Pump

Status Screen

PREPARING TO PRIME

Disconnect from Body

Insert/Lock Reservoir
Hold ACT to Prime

ESC

ACT

ACT

time out

(HOME screen)