Troubleshooting – Health O Meter PRO PLUS 2650KL User Manual

Page 20

Advertising
background image

19

TROUBLESHOOTING

Refer to the following instructions to check and correct any failure before contacting service
personnel.

SYMPTOM

POSSIBLE CAUSE

CORRECTIVE ACTION

Scale does not turn on

1. Dead Battery

2. Faulty electrical outlet

3. Bad power supply

1. Replace

batteries

2. Use a different outlet

3. Replace

adapter

Questionable weight or
the scale does not zero

1. External

object

interfering with the scale

1. Remove interfering object

from the scale

2. The display did not show

”0.0” before weighing

2. Ask the patient to get off

the scale, zero the scale
and begin weighing
process again

3. Scale is not placed on a

level floor

3. Place the scale on a level

floor and begin weighing
process again

4. Scale is out of calibration 4. Check weight with known

weight value

5. Improper Tare

5. Place the item to be tared

on the scale. Press
REWEIGH. Once the
weight of the item is
displayed, press TARE.
Place the patient and the
tared item back on the
scale. Press REWEIGH
again.

Weighing is performed but
the display shows “weigh”
and “reweigh” every few
seconds; the weighing
process takes too long and
no weight is displayed.

The patient is not sitting still

Ask the patient to be still or
you can change to live weight
setting

The display shows
“Overload” message

The load on the scale
exceeds the capacity
(1000 lbs/450 kgs)

Remove the excess weight
and use the scale according to
its limits

The display shows
“LOW BAT” message

The batteries are empty

Replace batteries according to
instructions

The display shows
“Load Cell Error” message

There is a problem with one
or more load cells or the load
cell cable is disconnected.

Check load cell cable
connection at the display and
platform assembly ports. If the
problem is not corrected, refer
to qualified service personnel
to replace the defective load
cell