White calibration for reflectives 22, English – Hasselblad Flextight 848 User Manual

Page 22

Advertising
background image

White Calibration for Reflectives

The Flextight scanner uses a white reference to make sure that all

elements in the scanner's CCD react consistently to light brightnesses.

The scanner has been calibrated from the factory, and the calibration

data is stored in the scanner’s flash PROM.

As the scanner ages, the properties of the light tubes may change

slightly. Therefore, you must redo the white calibration every six months

or so and immediately after changing the light tubes.

Use procedure below to make White Calibration:

1. Find the white calibration sheet that came with your scanner. It is a

square white piece of paper (219 x 219 mm (8.6" x 8.6")).

If you have lost the white calibration sheet, then you can use a flat

(non-textured), perfectly white piece of paper of the same

dimensions.

2. Your scanner came with several clear plastic layers for the reflective

original holder. If you have used the reflective holder before, make

sure that there are no finger prints or other marks on the plastic

layer. Replace the layer with a new one if necessary.

3. Align the top edge of the sheet with the dotted line near the top of

the original holder (see Figure 6 opposite). Align each side of the

target with the edges of the original holder.

4. Select White Calibration from the Maintenance window. The

External White Calibration

window appears.

User's Guide - Flextight 848

English

Hardware Reference - Flextight 848

22

Calibrating the Scanner

Advertising