3 - 2

Quick Check® 890 User’s Guide

4. For the PDA version, select Device and then Scan from the Scan Data tab.

Calibrating the QC890 Verifier

When calibrating the QC890 Verifier system, the QC890 Verifier must be

connected to the Quick Check ImageData software. During calibration, it is

important to align the center of the barcode symbol to the center arrows, and

keep the QC890 Verifier flat and level to the surface of the reflectance page.
The reflectance page is a precision standard that is made of photographic

materials. To ensure proper calibration, you must replace the reflectance page

if it becomes dirty or physically damaged. Make sure to use only the original

calibration tools that were supplied with your QC890 Verifier kit or certified

replacements obtained from Hand Held Products.
1. Place the verifier on top of the reflectance page. Use the symbol alignment

window to center the reflectance page.

2. Press the scan button on top of the unit to turn on illumination to aid in

alignment.

Note: The best calibration is performed by aligning the center of the symbol to

the center arrows.

3. Press the scan button on top of the unit a second time while the illumination

is on to scan the calibration barcode. You will hear a series of beeps and the

Calibration is a requirement of the ISO/IEC standards and must be

performed with each aperture change.

!