Calibrating an external pen/mouse wand, Calibrating an external pen/mouse wand -4, Mouse wand. see – HandHeld Entertainment Quick Check ImageData Barcode Verifier Quick Check 890 User Manual

Page 20

Advertising
background image

3 - 4

Quick Check® 890 User’s Guide

Hold the mouse wand horizontally so it moves across the barcode in the

direction of the arrows:

3. Lightly touch the wand against either quiet zone (clear area) on either side of

the barcode. For the pen wand, you will have the proper scanning angle

when the tip and the reflectance guide (black plastic clamp) are flat on the

scanning surface.

4. At about the same speed with which you would quickly underline a sentence

in this manual, and without interruption, gently move the wand across the

center of the barcode. You do not need to press hard on the wand while

scanning the barcode. Gentle pressure on the wand will not scratch the

contacted surface, but pressing down hard may damage the code and yield

poor results.

5. Listen for audible tones.

• If you hear a single beep tone followed by another single beep of the same

tone, it signifies a good read.

• If you hear a single beep and then a beep of a different tone, the unit

detected an error in the scanned symbol.

Calibrating an External Pen/Mouse Wand

1. Using the proper scan technique described in step 4 above, scan the

reflectance page. You will hear a single beep after the first good scan and

the following display appears:

2. Continue to scan the reflectance page for a total of 10 good scans. After

each good scan, a single beep sounds and the display indicates the number