Using quick check® pc, Using quick check – HandHeld Entertainment Quick Check PC600 User Manual

Page 28

Advertising
background image

Message Length

Right Click Msg. Length display

Select variable or fixed message length

Note: If fixed-length message format is selected, it will apply to all

symbologies enabled.

Wide-to-Narrow Ratio

Right Click W/N Ratio display

Enable/disable Wide-to-Narrow element ratio checking

Set W/N ratio requirements

Speed Variance

Right Click Speed Variance display

Enable/disable display of scanning speed variance

Traditional Measures are not user-definable. Values set in the software are
shown below.

Rl

≥ 25% minimum

Rd

≤ 30% maximum

PCS

≥ 75% minimum

Using Quick Check

®

PC

Symbols should be verified in their final form. That is, if a label is to be applied to
a corrugated container, the label should be affixed to a sample of the corrugated
material before verifying the symbol.

The requirement to verify a symbol in its final configuration is based on
experience. Thin label stock (high show-through), overlaminates, and dark or
patterned packaging material may significantly alter a symbol's quality from what
might be achieved if the symbol is verified in a virgin state under ideal conditions.

It is almost always best to verify the symbol in the final form. However, if you are
in doubt about a specific industry or customer requirement, check the standard
before proceeding.

Aperture: Hand Held Products offers mouse and pen wands with apertures
designed to match the most common X dimensions. Wands are available with
apertures of 0.003 (Mouse Wand only), 0.005, 0.006, 0.010, and 0.020 (Mouse
Wand only) inches. These aperture sizes correspond to those specified in
ISO/IEC 15416. The proper aperture should be selected based on the X
dimension range shown in the tablebelow.

20

Advertising