Table 4. Flatness

Test instrument

and measuring receiver

settings (MHz)

1

2
4
6
8

10
12
14
16
18
20

b. Adjustments. No adjustments can be made.

15. Square Wave Amplitude and Flatness

a. Performance Check

(1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50

Ω

feedthrough

termination.
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below:
(a) SIGNAL off (option 002 not lit).
(b) ENTRY DC OFFSET.
(c) 0 VOLT.
(d) FUNCTION square wave.
(e) ENTRY AMPTD.
(f) 5 V RMS.
(g) ENTRY FREQ.
(h) 100 Hz.
(i) AMPTD CAL.

(3) Multimeter will indicate between 4.95 and 5.05 V ac.
(4) Enter 1 kHz using DATA keys. Multimeter will indicate between 4.95 and 5.05
V ac.
(5) Enter 100 kHz using DATA keys. Multimeter will indicate between 4.95 and
5.05 V ac.