Table 5. Dc Offset - Continued

Test instrument

DC OFFSET

Multimeter indications

(V dc)

settings

Min

Max

14.99 mV

0.01493

0.01505

-14.99 mV

-0.01505

-0.01493

4.999 mV

0.004979

0.005019

-4.999 mV

-0.005019

-0.004979

1.499 mV

0.001479

0.001519

-1.499 mV

-0.001519

-0.001479

NOTE

Perform (8) through (11) below for option 002 only.

(8) Remove 50

Ω

feedthrough termination from equipment setup.

(9) Press SIGNAL key on.
(10) Enter 20 VOLT using DATA keys. Multimeter will indicate between 19.775
and 20.225 V dc.
(11) Enter -20 VOLT using DATA keys. Multimeter will indicate between -19.775
and -20.225 V dc.

b. Adjustments. No adjustments can be made.

18. Dc Offset with Ac

a. Performance Check

(1) Connect SIGNAL to multimeter INPUT HI and LO using 50

Ω

feedthrough

termination.
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below:
(a) SIGNAL off (option 002 not lit).
(b) FUNCTION sine wave.
(c) ENTRY FREQ.
(d) 20.999 999 999 MHz.
(e) ENTRY AMPTD.
(f) 1 VOLT.
(g) ENTRY DC OFFSET.
(h) 4.5 VOLT.
(i) AMPTD CAL.

(3) Multimeter will indicate between 4.35 and 4.65 V dc.