Para 5, Table 2 – HP 3325A User Manual

Page 5

Advertising
background image

TB 9-6625-2049-35

5

5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
50

feedthrough termination, Hewlett-Packard, Model 11048C.

Table 2. Minimum Specifications of Equipment Required

Common name

Minimum use specifications

Manufacturer and model

(part number)

AUDIO ANALYZER

Output: 4.5 V, 10 kHz
AM distortion:

-30 dB at 10 kHz Distortion:

-60 dB at 50 Hz

Boonton, Model 1120-S/10
(MIS-35954/2)

FREQUENCY COUNTER

Range: 99,9950

µ

s to 60,000,300 Hz

Accuracy: 1.25 x 10

-8

Capability: Time interval A to B

Hewlett-Packard, Model 5345A
(MIS-28754/1 Type 1)

MEASURING RECEIVER

Frequency: 100 kHz to 10 MHz
Volts: .683 to .732 V rms
Accuracy:

±

0.85%

AM: 0 to 80%

Consisting of : Measuring
receiver Hewlett-Packard,
Model 8902A and Sensor
module Hewlett-Packard, Model
11722A

MULTIMETER

Range: -20.225 to +20.225 V dc
Accuracy:

±

0.05%

Range: 0.3416 to 20.4 V ac
100 Hz to 100 kHz
Accuracy:

±

0.3%

Hewlett-Packard, Model
3458A

OSCILLOSCOPE

Frequency: 1 kHz to 10 MHz
Amplitude: 1.8 to 2.2 V p-p
Accuracy:

±

2.5%

Capabilities
Duty cycle: 49.7 to 50.3%
Overshoot: 5%
Rise time and fall time: <20 ns

Tektronix, Type 2430A
(OS-291/G)

SPECTRUM ANALYZER

Frequency: 50 kHz to 20 MHz
Input: -20 dBm and 40 V
Display capability: <-65 dBc

(AN/USM-489A(V)1)

SYNTHESIZER/LEVEL
GENERATOR

Frequency: 0.1 MHz
Amplitude: 13 dBm

Hewlett-Packard, Model
3335AOPT 001-K06
(MIS-35938)

TIME FREQUENCY
WORKSTATION

Reference output frequency: 10 MHz
Accuracy:

±

5 x 10

-8

Autek Systems Corporation,
Model 620 (MIS-38946)

RESISTANCE STANDARD

Range: 470

Biddle Gray, Model 71-631
(7910328)

Advertising