Battery recharging, Battery replacement – Fisher 1266-X User Manual

Page 20

Advertising
background image

18

BATTERY RECHARGING

Rechargeable nicad batteries may take as many as 1,000

recharges, however they do have some limitations you should

be aware of:

1. Nicads will last only about half as long as

standard carbon-zinc batteries before

recharging is required.

2. Nicads may develop a “memory” if you give

them repeated “booster” charges. In other

words, if you charge your batteries over and

over again for only three hours at a time, the

BATTERY REPLACEMENT

1. Two drop-in battery packs are located in the

battery housing below the armrest. Each pack

contains four 1.5 volt AA batteries.

2. Remove the battery end cap by partially

unscrewing the knob on the battery end cap.

3. Remove the battery packs and replace the

batteries. Make sure the batteries are oriented

correctly as indicated by the polarity marks on

the inside of the packs.

4. Reinstall the battery packs, connector end first.

Note that the large split connector with the

spring behind it is the negative terminal and the

smaller terminal is the positive terminal. Both

should match up with the polarity markings on

the battery end cap. Another way to tell if the

packs are in correctly is by checking the position

of the rivets on the back end of the packs: the

rivets should always be facing in towards the

center of the battery housing.

5. Put the battery end cap back on and lock it in

place by tightening the knob, finger tight only.

NOTE: If either or both of the battery packs are installed

incorrectly, the 1266-X will simply not turn on. The instrument

will not be damaged.

Advertising
This manual is related to the following products: