Displaying temperatures, Holding the displayed temperatures, Viewing the min, max, and avg readings – Fluke 54IIB User Manual

Page 18

Advertising
background image

53IIB/54IIB
Users Manual

14

Displaying Temperatures

1. Press

to select the correct temperature scale.

2. Hold or attach the thermocouple(s) to the

measurement location.

The temperature reading appears in the selected
display.

3. Model 54: Press

to toggle between showing the

T1, T2, and T1-T2 reading in the primary or
secondary display.

Notes

The display shows "- - - -" when a thermocouple
is not connected.

The display shows  (overload) when the
temperature being measured is outside the
thermocouple's valid range.

Model 54: If only thermocouple T2 is connected,
the T2 reading appears in the primary display.

Holding the Displayed Temperatures

1. Press

to freeze the readings on the display.

The display shows .

2. Model 54: Press

to toggle showing the T1, T2, or

T1-T2 readings in the primary or secondary display.

3. Press

again to turn off the HOLD function.

Viewing the MIN, MAX, and AVG Readings

1. Press

to step through the maximum (MAX),

minimum (MIN), or the average (AVG) readings.

The elapsed time since entering MIN MAX mode, or
the time at which the minimum or maximum occurred,
appears on the display.

2. Press

,

(CANCEL) to exit MIN MAX mode.