Making Measurements

13

The following are some tips for measuring voltage:

When you measure voltage, the meter acts
approximately like a 10 M

(10,000,000

)

impedance in parallel with the circuit. This loading
effect can cause measurement errors in high-
impedance circuits. In most cases, the error is
negligible (0.1% or less) if the circuit impedance is
10 k

(10,000

) or less.

For better accuracy when measuring the dc offset of
an ac voltage, measure the ac voltage first. Note the
ac voltage range, then manually select a dc voltage
range equal to or higher than the ac range. This
procedure improves the accuracy of the dc
measurement by ensuring that the input protection
circuits are not activated.

MIN MAX

RANGE

HOLD

H

Hz

REL

mA

A

mV

V

V

OFF

!

!

1000V MAX

400mA MAX

FUSED

10A MAX

FUSED

PEAK MIN MAX

µ

A

CAT II

V

87

TRUE RMS MULTIMETER

MIN MAX

RANGE

HOLD

H

Hz

REL

mA

A

mV

V

V

OFF

!

!

A

COM V

mA

µ

A

A

COM V

mA

µ

A

1000V MAX

400mA MAX

FUSED

10A MAX

FUSED

PEAK MIN MAX

41/2 DIGITS

1 Second

41/2 DIGITS

1 Second

µ

A

CAT II

Switch Box

V

+

AC Voltage

DC Voltage

III

87

TRUE RMS MULTIMETER

III

iy2f.eps

Figure 2. Measuring AC and DC Voltage