Ti10, Ti25, TiR and TiR1
Users Manual

18

Ti10............................................................5

°

C

Ti25, TiR, and TiR1 ....................................2.5

°

C

Minimum span (in auto mode)

Ti10............................................................10

°

C

Ti25, TiR, and TiR1 ....................................5

°

C

IR-Fusion

®

Information

Visual and IR Blending (Ti25 and TiR1 only)

Picture-In-Picture (PIP)

Ti10 and TiR ..............................................100 % IR displayed in center 320 x

240 pixels

Ti25 and TiR1.............................................Three levels of on-screen IR blending

displayed in center 320 x 240 pixels

Full Screen (PIP off)

Ti10 and TiR ..............................................100 % IR displayed on 640 x 480 LCD

Ti25 and TiR1.............................................Three levels of on-screen IR blending

displayed on 640 x 480 LCD

Ti25 and TiR1 allows user to adjust palette, alpha blend, level, span, PIP, and
emissivity on a captured image before it is stored.

Voice Annotation..................................................60 seconds maximum recording time

per image (TiR1 and Ti25 only)

Image and Data Storage

Storage Medium.................................................SD Memory Card (2 GB memory card

will store at least 1200 fully radiometric
(.is2) IR and linked visual images each
with 60 seconds voice annotations or
3000 basic (.bmp) IR images)

File Formats .......................................................Non-Radiometric (.bmp) or Fully-

Radiometric (.is2)
No analysis software required for Non-
Radiometric (.bmp) files

Export File Formats w/ SmartView

®

Software .....JPEG, JPG, JPE, JFIF, BMP, GIF,

DIP, PNG, TIF, and TIFF