Calibration Information

6

Required Test Equipment

Table 2 lists the required test equipment needed to perform the performance test and calibration procedures
for the 50S and 50D thermometers. If the required equipment is not available, instruments with equivalent
specifications may be used.

Table 2. Required Test Equipment

Test Equipment

Required Characteristics

Recommended Model

DMM

DC Voltage Accuracy: 0.5%

Fluke 87 Series IV Digital
Multimeter

Temperature Probe

Accuracy: Certified to

±

0.2

o

C ambient

Fluke Model 80T-150U

DC Voltage Calibrator

Output Voltage 0 to 10 V
Accuracy: 0.025%
Resolution: 10

µ

V

Fluke 5520A

Elastomeric Switch Pad

Fluke Part Number 680686

Thermocouple Connector

For K-type thermocouple

Fluke 80CK-M
Fluke Part Number 779942

Thermocouple Connector

For J-type thermocouple

80CJ-M
Fluke Part Number 833046

Thermocouple Wire
(K or J Type)

Standard

Omega

Dewar Flask/Cap

1-pint capacity

Thermos

Mercury Thermometer
(either

o

C or

o

F)

0.02

o

C resolution

0.05

o

F resolution

Princo Model ASTM56C
Princo Model ASTM56F

Performance Tests

Performance tests are recommended for incoming inspection, periodic maintenance, and for verifying the
thermometer’s specifications. If the thermometer fails any part of the performance tests, calibration and/or
repair is indicated.

Use the following procedure to prepare the thermometer before beginning the performance tests or
calibration procedures.

1.

Allow the thermometer to stabilize to room temperature, 23

o

C

±

5

o

C (73

o

F

±

9

o

F).

2.

Check the battery and replace it if necessary.

Display Test

Turn the thermometer on while pressing the HOLD button. All display segments are shown until the HOLD
button is released. Check that all the display segments are showing as in Figure 3.

Note

Because pressing the HOLD button during power-up activates one of the thermometers other
functions, take care to return the instrument to its normal function (if necessary).