Fluke 115 User Manual

Page 15

Advertising
background image

Digital Multimeters

Calibration Adjustment

15

Notes

Set the calibrator to Standby prior to changing the function switch position and after
completing adjustment of each function.

If the calibration adjustment procedure is not properly completed, the Meter will not operate
correctly.

Table 4. Calibration Adjustment Steps

Calibration Steps

Rotary

Switch

Position

114 115

[1]

116

[1]

117

[1,2]

Input

Terminals

Calibrator

Source Value

e

Ohms

N/A

C/01

C/01

C/01

[2]

No leads

No leads

C/01

C/02

C/02

C/02

V

Ω

/+ and COM

0 V, 0 Hz

C/02

C/03

C/03

C/03

V

Ω

/+ and COM

300 mV, 0 Hz

C/03

C/04

C/04

C/04

V

Ω

/+ and COM

100 mV, 0 Hz

C/04

C/05

C/05

C/05

V

Ω

/+ and COM

-300 mV, 0 Hz

C/05

C/06

C/06

C/06

V

Ω

/+ and COM

60 mV, 0 Hz

C/06

C/07

C/07

C/07

V

Ω

/+ and COM

600 mV, 0 Hz

m

C/07

C/08

C/08

C/08

V

Ω

/+ and COM

600 mV, 60 Hz

C/08

C/09

C/09

C/09

V

Ω

/+ and COM

600

e

, 2-wire comp

C/09

C/10

C/10

C/10

V

Ω

/+ and COM

6 k

e

C/10

C/11

C/11

C/11

V

Ω

/+ and COM

60 k

e

C/11

C/12

C/12

C/12

V

Ω

/+ and COM

600 k

e

C/12

C/13

C/13

C/13

V

Ω

/+ and COM

6 M

e

[3]

C/13

C/14

C/14

C/14

V

Ω

/+ and COM

Short

[3]

e

Ohms

C/14

C/15

C/15

C/15

V

Ω

/+ and COM

40 M

e

[3]

C/15

C/16

C/16

C/16

V

Ω

/+ and COM

6 V, 60 Hz

C/16

C/17

C/17

C/17

V

Ω

/+ and COM

60 V, 60 Hz

C/17

C/18

C/18

C/18

V

Ω

/+ and COM

600 V, 60 Hz

C/18

C/19

C/19

C/19

V

Ω

/+ and COM

6 V, 0 Hz

C/19

C/20

C/20

C/20

V

Ω

/+ and COM

60 V, 0 Hz

K

C/20

C/21

C/21

C/21

V

Ω

/+ and COM

600 V, 0 Hz

Set calibrator to standby, reconfigure leads, and program for amps output.

?

N/A

C/22

N/A

C/22

A and COM

6 A, 60 Hz

[3]

A

N/A

C/23

N/A

C/23

A and COM

6 A, 0 Hz

$

AC

μ

amps

N/A N/A

C/22

N/A

+ and COM

600

μ

A, 60 Hz