Testing the voltalert function (117 only) – Fluke 115 User Manual

Page 11

Advertising
background image

Digital Multimeters

Performance Tests

11

Table 3. DMM Performance Tests (cont)

Display Reading

Step Function Range Applied

114

[1]

115

[1]

116

[1]

117

[1]

41

VoltAlert

Hi

N/A N/A N/A

Refer to

steps 1 – 5

in the

procedure

below

42

VoltAlert

Lo

N/A N/A N/A

Refer to

steps 6 – 9

in the

procedure

below

[1] If using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency.

[2] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a 5 second delay between

increments.

Testing the VoltAlert Function (117 only)

Use the following procedure to verify that VoltAlert functions properly.

Note

Ensure the instrument is REMOVED from the holster prior to performing the test.

Keep the meter away from electrical noise sources during the tests, i.e., florescent

lights, dimmable lights, motors, etc.. These sources can trigger VoltAlert and
invalidate the test.

It may be necessary in steps 4 and 8 below to slightly adjust the Meter’s position for

maximum signal strength, in order to get the Meter’s beeper to sound continuously.

Refer to Figure 4 while performing the following steps.

1. Select the VoltAlert function, and verify that “Hi” is on the display. Verify that the beeper is silent and

the red LED is off.

2. Connect a double banana plug to the output voltage terminals of the calibrator (Fluke 5500A or

equivalent).

3. Set the calibrator output to 10 V at 60 Hz.

4. Hold the Meter so that the Meter’s top is vertically and horizontally centered and contacting the banana

plug’s Hi terminal. Verify that the Meter’s beeper is on continuously, and the red LED, at the top of the
display, lights up.

5. Place the calibrator in standby mode and verify that the beeper is now silent and the red LED is off.

6. Press the q(display should indicate ‘Lo’ range).

7. Set the calibrator’s output to 30 V at 60 Hz.

Advertising