Testing the display, Backlight test, Keypad test – Fluke 115 User Manual

Page 8: Preparing for the performance tests

Advertising
background image

114, 115, 116, and 117
Calibration Information

8

Table 2. Required Equipment (cont)

Recommended Equipment

Measurement Function

Accuracy

5500A Multi-product Calibrator (or
equivalent)

Frequency (115, 116, and 117)

2 V, 50 kHz

±

0.025 %

Fluke 80 AK
K-type Thermocouple Adapter Accessory

Temperature (116)

K-type Thermocouple, mini-plug on both
ends

Temperature (116)

Double Banana plug

VoltAlert (117)

Testing the Display

Push K and turn the rotary switch to the e position. Compare the display with the example in Figure 3.
Check all segments for clarity and contrast.

erc022f.emf

Figure 3. Display Segments

Backlight Test

To Test the Backlight, press the Q button and verify that the backlight comes on.

Keypad Test

To test the keypad, turn the Meter to ACV and push each button separately. Each button push should cause the
Meter to beep and activate a display annunciator.

Reset the Meter by turning it

Off

and then back to an On position.

Preparing for the Performance Tests

XW

Warning

To avoid possible electric shock or personal injury:

Do not perform the following procedures unless qualified to do so. Some
procedures involve the use of high voltages.

Before handling the test connections and in between tests, make sure
the calibrator is in standby mode (STBY).

To prepare for the performance test:

1. Make sure that you have the required equipment (refer to Table 2).

Advertising
This manual is related to the following products:

116, 117, 114