Fluke 561R User Manual

Page 2

Advertising
background image

Laser “On” symbol

SCAN or HOLD

ºC/ºF symbol (Celsius/Fahrenheit)

Primary temperature display

Secondary temperature display

Emissivity LO, MED, HI

Temperature values for the MIN, MAX, DIF, KTC.

KTC indicates the thermocouple temperature.

Low Battery symbol. Appears when the battery charge

is < 25%

Thermometer Display

Connecting the Temperature Probe

With probe inserted, the 561 stays on for 10 minutes (with laser

off) after trigger is released.

3

4

5

6

2

1

1

2

3

4

5

6

7

8

8

7

Locating a Hot and Cold Spot

Distance and Spot Size

Field of View

Switching between °C and °F

Switch

D:S=12:1

38 mm @

300 mm

75 mm @

900 mm

132 mm @

1500 mm

1.5 " @

12 "

3 " @

36 "

5.3 " @

60 "

S

D

Yes

No

Specifications

Infrared

Measurement Range............... -40 °C to 550 °C (-40 °F to 1022 °F)

Spectral Range........................ 8 to 14 microns

Accuracy................................. ±1 % or ±1 °C (2 °F); < 0 °C (32 °F),

±1 °C (2 °F) ±0.1°/1°

(Assumes ambient operating temperature of

23 to 25 °C (73 to 77 °F))

Repeatability ........................... ±0.5 % of reading or ±1°C (2 °F)

Display Resolution .................. 0.1 °C (0.1 °F)

Secondary Display Information Maximum, Minimum, Differential, KTC

Response Time (95 %) ........... 500 ms

Distance to Spot (D:S) ............ 12:1

Emissivity Adjustment............. Three settings: low (0.3), medium (0.7),

high (0.95)

Contact Probe Input

Input Temperature Range........ -40 °C to 550 °C (-40 °F to 1022 °F)

Input Accuracy........................ Input accuracy +/- 1.1 °C (+/- 2 °F)

Display Resolution .................. 1 °C (1 °F)

Laser

Sighting .................................. Single point laser

Power ..................................... Class 2 (II) operation; Output <1 mW,

wavelength 630 to 670 nm

Bead Thermocouple Probe

Type ........................................ Type K with miniconnector

Measurment Range................. -40 °C to 260 °C (-40 °F to 500 °F)

Accuracy................................. ±1.1 °C (2.0 °F) from 0 °C to 260 °C (32 °F to

500 °F). Typically within 1.1 °C

(2.0 °F) from -40 °C to 0 °C

(-40 °F to 32 °F)

Cable Length........................... 1 m (40 in) terminated with Type K

thermocouple beads

Electrical

Power Supply.......................... 2 AA Batteries (alkaline or NiCD)

Power Consumption ............... At least 12 hours battery life

Physical

Weight .................................... 0.322 kg (0.7099 lb)

Size ......................................... 17.69 cm (6.965 in) x 16.36 cm (6.441 in) x

5.18 cm (2.039 in)

Environmental

Operating Temperature Range 0 °C to 50 °C (32 °F to 120 °F)

Relative Humidity.................... 0 to 90 %, noncondensing up to 30 °C (86 °F)

Storage Temperature............... - 20 °C to 65 °C (-4 °F to 150 °F)

Optional Accessories............... Soft Case