4 configuration functions – Fluke 943-227-15 User Manual

Page 23

Advertising
background image

943-227-15 Digital Preamplifier
Operator Manual

3-5

For UDR systems, the address of the 942-200-80 Communications board in the UDR must be set 4040
and the baud rate must be set to 4800 baud. If the wiring of the Communications option board to the P2
connector is suspected, a loop test can be performed as described in "Loop Test Mode". The instruction
manual for the Communications option board (P/N 942-200-80-1) contains information on how to set up
the serial board in the UDR.

Firmware Version

This system requires a PROM for operation. For the part number and the latest revision of the PROM,
see the Factory Test Data Sheet. This Manual has been prepared for use with the P/N 227LNK, Rev. 2,
PROM.

3.4 Configuration Functions

The preamplifier configuration jumpers and required settings for this application are listed below:

Table 3-2. Preamplifier Serial Interface Board 977-210-10M Switch Settings

Switch Normal

Op.

Switch Function

SW1

Open

Microprocessor Reset switch, momentary contact

SW2

N/A

Not supplied

SW3-1 ON OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)
SW3-2

OFF

OFF • (No function is implemented in the 977-210-10M)

SW3-3

OFF

OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for

factory PGA and Auto Zero set-up only)

SW3-4

OFF

OFF • (No function is implemented in the 977-210-10M)


NOTE

SW3 is a 4-position DIP switch on the 977-210-10M serial
interface board.



Table 3-3. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M

Jumper Normal

Setting

Function Alternate

Setting

Function

JMP1

A to B

Enables

Victoreen Loop

(normal)

B to C

Enables RS232C

JMP2

N/A

Deleted, Not Used

N/A

Deleted, Not Used




Table 3-4. Jumper Positions for Preamplifier Electrometer Board 977-200-15M

Jumper Normal

Setting

Function Cal.

Position

Function

JMP1

A to B

Normal Operation

B to C

PGA Offset Adjust *


* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment



Advertising