943-227-15 Digital Preamplifier
Operator Manual

1-4

Table 1-3. Preamplifier Serial Interface Board 977-210-10M Switch Settings

Switch Normal

Op.

Switch Function

SW1

Open

Microprocessor Reset switch, momentary contact

SW2

N/A

Not supplied

SW3-1

ON

OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)

SW3-2

OFF

OFF • (No function is implemented in the 977-210-10M)

SW3-3

OFF

OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for

factory PGA and Auto Zero set-up only)

SW3-4

OFF

OFF • (No function is implemented in the 977-210-10M)


NOTE

SW3 is a 4-position DIP switch on the 977-210-10M serial
interface board.



Table 1-4. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M

Jumper

Normal Setting Function

Alternate Setting Function

JMP1

A to B

Enables

Victoreen Loop

(normal)

B to C

Enables RS232C

JMP2

N/A

Deleted, Not Used

N/A

Deleted, Not Used




Table 1-5. Jumper Positions for Preamplifier Electrometer Board 977-200-15M

Jumper

Normal Setting Function

Cal. Position Function

JMP1

A to B

Normal Operation

B to C

PGA Offset Adjust *


* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment