2 set-up – Fluke 943-227-15 User Manual

Page 20

Advertising
background image

943-227-15 Digital Preamplifier
Operator Manual

3-2

NOTE

The electrometer circuitry in the preamplifier is highly sensitive to
moisture and physical damage. Use extreme care when the
preamplifier case is opened as the high impedance electrometer
circuitry is easily damaged. Do not leave the preamplifier
enclosure in the open position for any extended period of time.

The preamplifier terminations are listed below.

Table 3-1. 943-227-15 Preamplifier Electrical Connections

943-227-15

943-27-15VL

TB1- Description

Connects To

977-210-10M

Description Connects To

977-210-10M

1 XMIT J4-1

+ VL, 15

VDC

J3-1

2 DTR

J4-2

+ TX

J3-2

3 GND

J4-3

- TX

J3-3

4 REC

J4-4

- VL

J3-4

5

Not Used

Not Used

+ RX

J3-5

6

Not Used

Not Used

- RX

J3-6

7

+ 15 VDC

J3-7

Not Used

Not Used

8 Ground J3-8

Ground

J3-8

9

- 15 VDC

J3-4

Not Used

Not Used

10

Not Used

Not Used

Not Used

J3-11

Detector

Connections

Detector

Connections

HV Detector

MHV

From Field

coaxial cable

Detector

MHV

From Field

coaxial cable

SIG Detector

BNC

Factory

Terminated,

To 977-200-

15M, R41

Detector

BNC

Factory

Terminated,

To 977-200-

15M, R41

3.2 Set-up

To place the system in operation, the following steps should be performed:

NOTE

Ensure you have read and fully understand
Section 1, 2 and 3 prior to continuing.

1. Verify that jumpers and DIP switches and jumpers on the preamplifier, are set for the operational

features desired. Refer to Section 1, Table 1-4.

2. Refer to the applicable controller manual prior to powering up the controller.

Advertising