Fluke 1595A User Manual

Page 66

Advertising
background image

1594A/1595A Super-Thermometer

Main Menu

56

In the Calibration Parameters screen the user may view and change measurement calibration settings. The

calibration parameters can be viewed with no password but a password is required to make changes.
Calibration parameters include calibrated values for each of the internal resistors and the linearity coefficients

for the resistance ratio measurements. Since the accuracy of the Super-Thermometer depends on these values,

they must be accurate. Refer to Section 9.4.2, ADJUST PARAMETERS (F2), on page 106 for instructions

needed to adjust the calibration parameters.
Apart from the calibration parameters, the Calibration Parameters screen also includes the following configu-

ration fields:
CALIBRATION DATE- used to record the date the Super-Thermometer is calibrated.
SI UNIT LOCK- enables or disables °F as an allowed temperature unit option. When this is set to ON, tem-

peratures may be displayed in units of °C and K only, and °F will not be available as an option for the Unit

setting (see Section 5.7.4.5, TEMPERATURE UNIT (F5), on page 50).