Probe Replacement

The modular probe of the food safety thermometer is replaceable. To
replace probe, extend the probe partially to access the rubber caps.
Carefully pry up the rubber caps using a pin, and remove the screws using
a #2 cross-tip screwdriver as shown. Grasp the probe and carefully
separate the old probe from the probe base. Install the new probe in the
reverse order, firmly tightening the screws and replacing the rubber plugs.
The calibration accuracy of the unit is not affected by probe replacement
(Figure 14).

ebq17.eps

Figure 14. Probe Replacement

Battery Insertion and Replacement

To install a new 9V battery, remove the rubberized battery compartment
“plug” at the base of the unit by grasping the sides of the cover and pulling
straight out, exposing the battery. Gently shake or tap the base of the unit
on your palm to gain access to the battery. The 9V battery of the unit is
connected to the polarized snap connector (Figure 15).

9V

ebq18.eps

Figure 15. Battery Insertion and Replacement