Changing to the J-Type Thermocouple

Press

HOLD

during power-up.

Changing to Low Resolution

To select the low display resolution [1.0

o

C (1.0

o

F)]:

1. Turn the thermometer off.

2. Press

T1

+

ON/OFF for 2 to 3 seconds.

The alternate resolution remains set until the
thermometer is turned off.

Adjusting the Offset

To compensate for thermocouple errors:

1. Plug the thermocouple into the input connector.

2. Place the thermocouple in a known, stable temperature

environment (such as an ice bath or a dry well
calibrator) and allow the readings to stabilize.

3. Simultaneously push and turn the OFFSET control until

the display shows the correct reading.

This adjustment affects calibration.