Ohms inputs – Fluke 2625A User Manual

Page 161

Advertising
background image

Appendices

Specifications

A

A-9

Ohms Inputs

Range Resolution

Typical Full

Maximum

Current

Through

Maximum Open

Slow Fast

Scale Voltage Unknown Circuit Voltage

300

Ω

10 m

Ω

0.1

Ω

0.22 V 1 mA 3.2V

3 k

Ω

0.1

Ω

1

Ω

0.25V 110

µ

A 1.5 V

30 k

Ω

1

Ω

10

Ω

0.29 V 13

µ

A 1.5 V

300 k

Ω

10

Ω

100

Ω

0.68 V 3.2 .

µ

A 3.2 V

3 M

Ω

100

Ω

1 k

Ω

2.25 V 3.2

µ

A 3.2 V

10 M

Ω

1 k

Ω

10 k

Ω

2.72 V 3.2

µ

A 3.2 V

4-Wire Accuracy ±(%±

Ω

)

Range 18°C to 28°C 0°C to 60°C

90 Days, Slow 1 Year, Slow 1 Year, Fast 1 Year, Slow 1 Year, Fast

300

Ω

0.013% + 20 m

Ω

0.014% + 20 m

Ω

0.014% + 0.2

Ω

0.031% + 20 m

Ω

0.031% + 0.2

Ω

3 k

Ω

0.015% + 0.2

Ω

0.016% + 0.2

Ω

0.016% + 2

Ω

0.039% + 0.2

Ω

0.039% + 2

Ω

30 k

Ω

0.013% + 2

Ω

0.014% + 2

Ω

0.014% + 20

Ω

0.039% + 2

Ω

0.039% + 20

Ω

300 k

Ω

0.020% + 20

Ω

0.021% + 20

Ω

0.021% + 200

Ω

0.050% + 20

Ω

0.050% + 200

Ω

3 M

Ω

0.059% + 200

Ω

0.063% + 200

Ω

0.063% + 2 k

Ω

0.231% + 200

Ω

0.231% + 2 k

Ω

10 M

Ω

0.168% + 2 k

Ω

0.169% + 2 k

Ω

0.709% + 20 k

Ω

0.573% + 2 k

Ω

0.923% + 20 k

Ω

2-Wire Accuracy

Add 4.0

Ω

to accuracy specifications for channels 1 to 20, and add 20 m

Ω

for channel 0.

Lead wire resistances are not included.

Input Protection

300V dc or ac rms on all ranges

Cross-Talk Rejection

Refer to Appendix D.

Advertising