Min, Max, Avg, and Differential readings are saved as part of the

saved data when either Min/Max or Avg/Dif mode is enabled.

Alarm

The thermometers have a programmable high and low alarm to designate high or

low readings depending on the thresholds entered. When the alarm level is

reached, an alarm will sound and the display will flash orange and white. To set

either the high or low alarm:
1. Press the Menu softkey until Alarm appears as the right softkey function.
2. Press the Alarm softkey to access the Alarm menu.
3. Press either the Hi or Lo softkey, depending on which alarm will be set.
4. Press the ON softkey to turn the alarm on.
5. Press the OFF softkey to turn the alarm off.
6. Use the Set softkey to access the Hi or Lo Alarm Set menu.
7. Use the down or up softkeys to change the alarm setting.
8. Once the desired settings have been entered, press the Done softkey.

Trigger Lock

The thermometer trigger can be locked on for continuous measurement. To lock

the trigger:
1. Press the Menu softkey until the lock symbol () appears as the left softkey

function.

2. Press the  softkey to lock the trigger. The lock symbol appears on the

display to designate a locked trigger. When the trigger is locked, the 

softkey changes to . Press this softkey to unlock the trigger.

Laser

*W

Warning

To prevent eye damage and personal injury:

Do not look into the laser. Do not point laser directly at

persons or animals or indirectly off reflective surfaces.

The thermometer is equipped with a laser used for aiming purposes only. The

laser turns off when the trigger is released.
To enable or disable the laser:
1. Press the Menu softkey until Laser appears as the right softkey function.
2. Press the Laser softkey to enable or disable the laser. * appears on the

display when the laser is enabled.