Maintenance

The calibration instrument has been designed with the utmost care. Ease of op-
eration and simplicity of maintenance have been a central theme in the product
development. Therefore, with proper care the instrument should require very
little maintenance. Avoid operating the instrument in an oily, wet, dirty, or
dusty environment.

• If the outside of the instrument becomes soiled, it may be wiped clean

with a damp cloth and mild detergent. Do not use harsh chemicals on the
surface which may damage the paint.

• Be sure that the well of the furnace is kept clean and clear of any foreign

matter. Do not use fluids to clean out the well.

• If a hazardous material is spilt on or inside the equipment, the user is re-

sponsible for taking the appropriate decontamination steps as outlined by
the national safety council with respect to the material.

• If the mains supply cord becomes damaged, replace it with a cord with

the appropriate gauge wire for the current of the instrument. If there are
any questions, call Hart Scientific Customer Service for more informa-
tion.

• Before using any cleaning or decontamination method except those rec-

ommended by Hart, users should check with Hart Scientific Customer
Service to be sure that the proposed method will not damage the equip-
ment.

• If the instrument is used in a manner not in accordance with the equip-

ment design, the operation of the furnace may be impaired or safety haz-
ards may arise.

• The over-temperature cut-out should be checked every 6 months to ensure

that it is working properly. Set the unit to 300°C and let it stabilize. Turn
the adjustable cutout knob down until the cutout is activated. Turn the
knob back up and push the reset button.

• Periodically remove the equilibration block and use emery cloth to re-

move the oxidation build up on the block.